Your First Clearance Is a Business Strategy! Not a Regulatory Task
Receiving a first regulatory clearance is one of the most critical milestones for any MedTech start-up. It is not the end of the journey, but rather the beginning of the next phase. Formally, clearance simply means that a company can finally market and sell its product. Yet for a start-up, it represents something even more significant: a major value inflection point. Clearance removes considerable risk and uncertainty from the proposition, making it an ideal moment to raise additional funds and unlock access to investors who only engage once a product has reached the market.
Achieving this first clearance with the limited resources typical of an early‑stage company is an intricate challenge that requires careful focus and prioritization. Ideally, the product, clinical, and regulatory strategies are aligned and optimized early in the process, enabling informed choices with the highest likelihood of success. In this article, we discuss three cases based on our work with clients. Certain details have been adjusted to allow a frank discussion of strategic trade-offs.
CASE I: Bioresorbable Implantable — FDA 510(k) vs EU-MDR Class III
As we discussed in last week’s article, choosing whether to pursue FDA clearance or EU‑MDR CE‑conformity assessment first can dramatically alter timelines, especially when the device classification differs between the two systems. One client had developed a novel biodegradable implantable device defined by a unique geometric shape intended to improve performance. The academic group behind the original concept had generated initial in‑human data with an earlier version of the device, and those results were promising. However, the prototype had not been developed under proper design controls, and it was sterilized using a completely different method, making it difficult to argue that the data reflected performance of the final design.
Under EU‑MDR Annex VIII Rule 8, the device would be classified as Class III because it is “wholly or mainly absorbed,” placing it in the highest risk category. This classification requires extensive evidence, including clinical investigations using a (near) final device, as well as notified body review and potentially additional scrutiny from EU expert panels. Although this pathway would allow the company to make claims about the performance benefits tied to the device’s geometry, it would also involve a lengthy, expensive, and high‑risk process.
The U.S. classification, by contrast, is not rule-based but relies on existing product codes. A search identified one product code that matched the intended use perfectly and categorized the device as Class II without special controls; a far lower regulatory burden. Initially, it appeared that the biodegradable device would not qualify because the code’s description referenced silicone rubber implants. However, reviewing the history of cleared devices revealed that, in the late 1990s, a manufacturer had successfully argued that its biodegradable design was substantially equivalent to silicone devices. Recent clearances in this code relied solely on preclinical performance and biocompatibility testing, with no human data required.
For this client, the U.S. route clearly provided a much lower barrier to initial clearance. By focusing first on the FDA pathway, the company could avoid redoing the clinical study for CE‑marking and potentially save years and millions of euros. The trade‑off, however, is that marketing claims in the U.S. would need to be more conservative. Additional clinical evidence would still be important for commercial uptake but collecting such data post‑market or outside the regulatory critical path significantly reduces business risk.
CASE II: Optimizing Predicate Selection in 510(k)
Even when the 510(k) pathway is appropriate, selecting the right predicate can save months of additional testing. Under EU‑MDR, Annex I defines the General Safety and Performance Requirements applicable to all devices, forming the basis for required testing and standards. Article 61 further mandates that a clinical evaluation must always be part of the development process, often requiring at least some pre‑CE‑marking clinical investigation.
In the U.S., the FDA uses the predicate device as the starting point for determining required testing. The logic is straightforward: if a new device is substantially equivalent to its predicate, the types of evidence used for the predicate should also be sufficient for the new device. This gives manufacturers strategic flexibility, as selecting a predicate with lower testing history can reduce evidentiary requirements.
For several clients, multiple potential predicates existed with comparable intended use and technical characteristics. In those situations, we recommended choosing predicates that did not include usability testing, even if the client had completed such testing for CE‑marking. FDA guidance requires usability studies to include at least 15 participants per user group, and offering non‑U.S. usability results requires a detailed justification of relevance to U.S. clinical practice. In many cases, this effectively triggers new U.S.-based usability testing, adding significant costs and delays.
The same strategic considerations apply to clinical evidence. FDA’s draft guidance states that if the predicate did not require clinical data, then clinical data are generally not required for the new device unless it falls within four narrow exceptions. Therefore, it is advantageous to select a predicate cleared without clinical evidence. Even when all predicates include clinical data, their study designs can differ significantly, making it worthwhile to analyse endpoints and sample sizes to minimize the required supporting evidence.
CASE III: Leveraging EU‑MDR Flexibility for a Novel System
For another client developing a system combining hardware with embedded software and non‑sterile disposables for premature infants, an extensive FDA database search revealed no predicate with a similar intended use. Since the device acts as adjunct to the standard of care the risk profile of the device is rather low even though it does address a serious condition in a vulnerable population. Initial clinical evidence showed positive effects that could support benefit claims.
Because no predicate existed, the U.S. pathway would require a De Novo submission. While successful De Novo reviews typically take 7–14 months, more than half of submissions are ultimately unsuccessful, making this a high‑risk and uncertain route for a start‑up.
However, under EU‑MDR the device components could be separated. The main active device could be classified as Class IIa, while the non‑sterile disposables could be considered accessories and classified as Class I. This approach means that only the Class IIa file would require notified body review. Because the active device does not contact the patient, a biological evaluation would not be necessary, reducing the number of notified body experts involved and lowering review costs. Additionally, shelf‑life testing for the disposables would not need to be completed before submission. The overall system performance would still be evaluated as part of the review, but the regulatory burden becomes significantly more manageable.
Conclusion
These cases demonstrate that minimizing time and effort to achieve initial market clearance requires looking beyond nominal review timelines. Greater savings are often found in the strategic selection of pathways, classifications, predicates, and evidentiary requirements. In the U.S., early engagement through the FDA’s Q‑submission process enables manufacturers to validate predicate choices and confirm evidence expectations, reducing regulatory uncertainty and business risk.
It is important to recognize that achieving initial regulatory clearance, whether in Europe or the US, does not guarantee commercial success. Market clearance does not equal market acceptance. At MD squared B.V. we work closely with founders to explore these strategic choices and have managed over 30 FDA Q‑submissions, consistently aiming to maximize the value of these interactions in reducing regulatory and business risks.