Why Regulatory Cliffs Distort Innovation
How step changes in regulatory burden drive design decisions instead of patient needs
In principle, most medical device regulatory frameworks are risk-based. This means that as the risks of a device increase, the amount of required evidence and thus the regulatory burden, should also increase. This sounds reasonable in theory. However, if the relationship between risk and burden is not monotonic but instead contains clear step-ups, there will be a strong incentive to stay just below those thresholds.
This is more than a theoretical concern. These “regulatory cliffs” directly shape product design decisions, intended use claims, and ultimately company strategy. Instead of optimizing for clinical value or patient outcomes, companies may optimize for regulatory positioning—adjusting features, indications, or technical approaches simply to avoid a step change in regulatory burden. This can result in sub-optimal products, non-conformities, delayed innovation, or even increased patient risk.
We discuss how the US FDA system and the EU MDR system differ in this respect, particularly regarding product classification, and why this can lead to significant and sometimes unexpected differences in regulatory burden. These differences materially impact development timelines, evidence strategies, and market access decisions. Three examples are provided that contrast these situations and illustrate the real-world impact of regulatory cliffs (note that these cases are based on publicly available information and are not clients of ours).
Regulatory frameworks
Both the FDA and the EU MDR define different classes of devices, in principle based on the risks they pose. With increasing classification, regulatory oversight increases. At first glance, the systems appear similar: the FDA defines Class I, II, and III, while the EU MDR defines Class I, IIa, IIb, and III. However, how devices are assigned to these categories—and the step changes in requirements between them—differs significantly. As a result, the relationship between risk and regulatory burden can be quite different in the two systems.
The EU MDR approach
Under the EU MDR, device classification is based on 22 rules defined in Annex VIII. These rules consider factors such as invasiveness (ranging from non-invasive, to invasive via body cavity, to surgically invasive, to implantable), duration of use for invasive devices, and type of energy for active devices. In general, these rules do a good job of producing classifications that intuitively make sense. Even for novel devices, they can usually be applied to determine an appropriate classification.
The type of evidence required for regulatory approval (e.g., biocompatibility, electrical safety, usability, risk management) is determined by which requirements from Annex I (General Safety and Performance Requirements, or GSPR) apply. Manufacturers must review the full list, justify which requirements are applicable, and explain how they will meet them. They must also justify the type and amount of evidence needed, taking into account the device’s intended purpose and risk profile.
Regardless of classification, a Clinical Evaluation must be performed, comparing the device’s risk-benefit profile with alternatives on the market and the current state of the art. While the amount and quality of evidence scale with the device’s risk, a formal risk-benefit determination is always required.
Similarly, regardless of classification, the manufacturer must implement a full Quality Management System (QMS) and compile a technical file for each product (or product group).
The largest step in regulatory burden is between Class I and Class IIa (and above). Manufacturers of Class I devices can self-certify: once they determine they meet the requirements, they can issue a Declaration of Conformity and place the product on the market. For Class IIa and higher, a Notified Body must be involved. This includes auditing the QMS and reviewing a representative set of technical files. After the initial conformity assessment, the company may place its devices on the market, but the Notified Body will conduct annual surveillance audits, evaluating both the QMS and any product changes.
Overall, this system means that the type and quality of required evidence can scale relatively smoothly—and largely monotonically—with the device’s risk profile.
The US FDA approach
In the US, the definition of device classification is quite different, and this is particularly problematic for novel devices. Formally, the first step is to identify the applicable regulation or product category based on the intended use. There are over 7,000 different product codes, often with very specific descriptions. If a device can be fit (“shoehorned”) into one of these categories, this determines its classification and, especially for more recent categories, the required evidence (special controls).
In practice, for Class II devices, the process typically starts with identifying a predicate device already on the market with the same intended use and similar technology. If the FDA accepts the predicate, the new device can follow the 510(k) pathway, requiring only a demonstration of substantial equivalence. This is usually done using similar tests, standards, and evidence as the predicate.
If no suitable predicate exists, the device is, by default, placed in Class III, requiring premarket approval (PMA), a much more involved pathway that typically includes extensive clinical evidence to demonstrate safety and effectiveness.
Since 1997, the FDA has also offered the De Novo pathway for devices that appear to be lower risk and may be appropriate for Class I or II. However, De Novo submissions still require substantial evidence (often including clinical data) to demonstrate a favorable risk-benefit profile. This pathway typically takes significantly longer than the 510(k) process (often over a year) and has a success rate of less than 50%. As a result, there is a strong incentive for companies to adjust their product description and intended use, and to argue that their novel device with new technology still fits within the scope of an existing predicate without introducing new risks.
Examples of regulatory cliffs

An example of a novel product being fitted into an ill-suited product code is the Da Vinci surgical robot. Its initial FDA clearance was obtained via the 510(k) pathway, claiming substantial equivalence to three predicate devices under the general “Endoscope and accessories” product category (GCJ). Because this product code primarily covered endoscope cameras and trocars, the applicable special controls and standards were not well suited to the system. As a result, the clearance required no clinical data, no usability study, and no formal determination of risk-benefit relative to alternative treatments or the state of the art. Under the EU MDR, this device is classified as Class IIb and would require a clinical evaluation critically reviewing the available clinical evidence for all claimed indications.
Conversely, when a device follows the De Novo pathway, the burden is high regardless of actual risk. The Sunrise Sleep Disorder Diagnostic Aid (SDDA) provides an instructive example. This device consists of a small patch worn on the chin during sleep, measuring mandibular (lower jaw) movement, with cloud-based software analyzing the data to detect obstructive sleep apnea. Because no suitable predicate was identified, it was cleared via the De Novo process.This required not only bench and biocompatibility testing, but also an extensive usability study and three clinical studies. In the EU, this device is classified as Class IIa, where similar types of evidence are required, but at a level closely aligned with its risk profile.

In the EU, by contrast, the main incentive to reduce classification relates to Class I devices, since manufacturers can self-certify without prior review. This has led to creative regulatory positioning, particularly for software. For example, some companies marketing AI-based medical scribes have argued for Class I classification to avoid Notified Body involvement. However, market surveillance actions, such as a recent inspectionby the Swedish competent authority, have challenged this approach, concluding that such devices are at least Class IIa and therefore require Notified Body review.
Conclusion
In the US, regulatory burden is often not directly aligned with device risk. Instead, the type of evidence required is largely driven by the identification of a suitable predicate. For novel devices, this creates a strong incentive to adjust indications for use or downplay technical differences to fit within an existing category. As a result, significant time and effort are spent arguing for predicate equivalence, rather than engaging in an open discussion about the appropriate type and level of evidence.
In contrast, under the EU MDR, there are fewer abrupt step changes in evidence requirements beyond the Class I boundary. This allows companies to define indications for use that provide the most business value, without being constrained by existing product categories. All relevant evidence types (e.g., usability, software testing, biocompatibility, risk management, and clinical evaluation) must be considered, and manufacturers must justify both the type and level of evidence based on the device’s intended purpose and risk profile. This also allows novel devices to adopt a more appropriate and tailored evidence strategy.
These differences between the US and EU frameworks, both in classification and in broader regulatory philosophy (see related article), are important when defining a company’s regulatory strategy. They affect not only initial market access but also long-term regulatory burden. At MD squared B.V., we work closely with our clients to develop optimal strategies for both regions. By explicitly considering regulatory cliffs, we help clients make informed and pragmatic choices.