Once You’re In: How EU MDR Can Enable Faster Innovation Than the FDA - If You Choose your NoBo Wisely
The EU Medical Device Regulation (EU‑MDR) is widely portrayed as a brake on MedTech innovation, especially when compared to the perceived speed and simplicity of the US FDA. Many founders and investors instinctively conclude that Europe is slow, rigid, and hostile to iteration, particularly for software medical devices.
This narrative is at least incomplete and in many cases, just wrong.
Once a company clears initial market access, the EU‑MDR can enable faster and more flexible innovation cycles than the FDA, especially for software as a medical device (SaMD). The catch is that this speed is not automatic. Under the EU‑MDR, post‑market innovation velocity is largely determined by founder choices: how the quality management system is designed, how the initial scope is defined, and crucially which notified body is selected.
In contrast, the US FDA follows a product‑centric regulatory model. Many software changes that are routine in development, such as algorithm improvements, indication updates, or expanded functionality, frequently trigger new pre‑market submissions. While mechanisms such as the Predetermined Change Control Plan (PCCP) offer some relief, innovation speed in the US remains tightly coupled to repeated FDA review.
In this article, we first explain the fundamental philosophical differences between the EU and US regulatory models: quality‑system‑driven oversight versus product‑level review. We then show how these differences translate directly into timelines for product updates and portfolio expansion. A critical (and often underestimated) factor in Europe is the variability in how notified bodies certify product ranges and assess significant changes. Drawing on our recent analysis of publicly available certification data, we illustrate why this variability matters.
We conclude with the strategic implications for MedTech founders and leadership teams—and the key decisions that determine whether EU‑MDR becomes a bottleneck, or a long‑term accelerator of innovation.
Difference in EU vs US regulatory philosophy: Quality management vs product focus
The EU‑MDR and the US FDA take fundamentally different approaches to the assessment of most medium‑risk medical devices (EU Class IIa and IIb non‑implantable devices and US Class II devices; see notes at the end of this article).

The EU approach: quality management as the foundation
Under the EU‑MDR, the starting point for market access is the manufacturer’s quality management system (QMS). In practice, companies typically implement their QMS in accordance with ISO 13485, which has been harmonized with EU‑MDR requirements.
Before a manufacturer can place a device on the EU market, a notified body performs a multi‑day audit of the company’s QMS. This audit covers both ISO 13485 requirements and the additional obligations introduced by the MDR. In parallel, the notified body conducts a detailed technical documentation review of at least one representative device per device category (for Class IIa devices) or per generic device (for Class IIb non‑implantable devices).
For start‑ups bringing their first device to market, this means that the full technical file of the first product is reviewed in detail. This process is expensive (often exceeding EUR 50,000) and time‑consuming (typically at least 6 to 12 months), and it is a major source of complaints about EU regulatory timelines.
Once both the QMS audit and the technical documentation review are successfully completed, the manufacturer receives a certificate from the notified body. This certificate is often informally referred to as a “CE certificate,” but this is a misnomer. In reality, it is a certificate confirming conformity of the QMS for a defined device range. A typical statement reads:
“The Notified Body hereby declares that the requirements of Annex IX, Chapter I of Regulation (EU) 2017/745 have been met for the listed products. The above‑named manufacturer has established and applies a quality management system, which is subject to periodic surveillance … The requirements of Annex IX, Chapter III are fulfilled.”
Importantly, this certificate confirms that the QMS is suitable for a defined range or category of products and that it will be subject to ongoing surveillance. The way in which the device range is described varies significantly between notified bodies, as discussed in a previous article. However, it should generally refer to device categories = MDA/MDN code for Class IIa or generic devices = EMDN code for Class IIb, not to individual products. As a result, certificates often contain broad descriptions such as “non‑active non‑implantable instruments” or “radiology instruments – medical device software.”
The US approach: product‑level review
The FDA takes a markedly different approach. For medium‑risk (Class II) devices, manufacturers must submit a separate premarket submission for each individual product they intend to market.
If a substantially equivalent “predicate device” exists, the 510(k) pathway can be used. Formally, FDA review takes 90 days and costs under USD 10,000 for small companies, which forms the basis for the perception that FDA pathways are faster and simpler.
However, as shown in a previous article, actual review times are typically longer. The median FDA review time for 510(k) submissions is around 167 days, largely because the FDA frequently raises additional questions that manufacturers must address. For novel devices without an appropriate predicate, the De Novo pathway is required, which is both longer and riskier. Reported success rates are around 40%, with a median duration of approximately 320 days.
Unlike the EU system, the FDA does not routinely audit or certify a company’s QMS as part of the 510(k) or De Novo process. While manufacturers selling into the US are required to implement a QMS (and as of February this year this is aligned with ISO 13485), there is no requirement for prior external certification. The FDA typically evaluates QMS compliance only during inspections.
Impact on design changes and the introduction of new products
These philosophical differences have major consequences for how regulatory changes are handled after initial market access.

Design changes under the EU‑MDR
Under the EU‑MDR (Annex IX, section 2.4), manufacturers must inform their notified body of any planned substantial changes to the QMS or the device range covered by the certificate. The notified body must then assess the proposed changes and determine whether additional audits are required to confirm continued compliance.
In principle, as long as updates or new products fall within the already certified device range, no prior review of the updated or new technical documentation is required. Instead, changes are assessed retrospectively as part of routine surveillance activities, including annual on‑site audits (Annex IX, section 3.3) and periodic sampling reviews of technical documentation (Annex IX, section 3.5).
In practice, this means that when a manufacturer plans updates, such as algorithm changes, indication updates, or the introduction of product variants within the approved range, they notify the notified body. The notified body typically reviews the notification within a few weeks and, in most cases, should not require pre‑approval. Once validation and verification are completed and the technical documentation is updated, the product can be released without delaying market access.
Design changes under the FDA
For FDA‑cleared Class II devices, the FDA provides guidance on when changes to an existing device require a new 510(k). This list is broad and includes many changes that are common in software development. For example, performance‑improving algorithm updates, changes to indications for use, or removal of contraindications generally require a new 510(k) submission. While formally reviewed within 90 days, these submissions often take considerably longer in practice.
Recognizing these challenges (particularly for AI/ML‑based devices) the FDA has introduced the concept of a Predetermined Change Control Plan (PCCP), (see guidance for AI and other medical devices). A PCCP allows manufacturers to implement predefined changes without submitting a new 510(k), provided those changes, and their validation methods, were approved as part of the original submission. However, PCCPs are inherently limited: any change outside the predefined scope still requires a new submission, and changes to indications for use are typically excluded.
Importantly, adding entirely new products to a portfolio in the US always requires a new 510(k) or De Novo submission.
Conclusions
For products that undergo frequent updates and iterations, such as software‑based devices, the EU‑MDR’s emphasis on a robust QMS combined with periodic surveillance can enable significantly faster innovation cycles than the FDA framework. In contrast, the FDA typically requires premarket review of many product changes, although mechanisms such as the PCCP can partially mitigate this burden.
A key challenge in the EU is that real‑world outcomes depend heavily on how individual notified bodies interpret the MDR and associated guidance. As discussed previously, there are notable differences in how notified bodies define product ranges on certificates and how they assess significant changes. For manufacturers selecting a notified body, it is therefore critical to understand their approach to evaluating changes, as this can have a substantial impact on future timelines and costs.
The large discrepancy between the regulatory effort required for initial market access and for subsequent updates also affects submission strategy. The concept of a “Minimal Regulatory Clearable Product” becomes particularly important. For example, we recently supported a client whose initial CE submission did not include all features required for full commercial rollout. Additional features were developed during the technical documentation review and could be released rapidly after initial clearance because they did not constitute substantial changes requiring prior notified body review.
At MD squared, we work closely with our clients to explore these trade‑offs and to identify strategies that optimize both time to market as well as market acceptance. This requires open and informed discussions, enabling clients to make informed choices aligned with their innovation, business and funding strategy.
Notes
Note 1 To streamline the discussion, this article focuses on medium‑risk devices: EU Class IIa and IIb (non‑implantable) and US Class II. EU Class IIb implantable and EU/US Class III devices are excluded. While similar principles apply, important differences warrant separate discussion. For low‑risk Class I devices, the EU and US approaches are again similar, as no proactive premarket assessment is required.
Note 2 As shown in a previous article, approximately 90% of devices follow the Annex IX conformity assessment pathway, requiring a full QMS including design and development. This article focuses on that pathway. Devices assessed under Annex XI may require a more limited QMS but at the cost of reduced flexibility for design changes.
Note 3 There is significant confusion around the definition of “substantial changes” as defined in section 2.4 of annex IX. There is no definition in the EU-MDR and also no relevant MDCG guidance. Notified bodies point still to the old NBOG 2014-3. The problem is that this only discusses the old MDD. This leads to a lot of confusion. The relevant section of the NBOG is chapter 5: “Criteria for “substantial” Changes”. However this covers 3 different sections each of which shows for which clause under the old directives this is deemed applicable. For Class IIa and IIb (non-implantable) the old Directive 93/42/EEC is the predecessor of the EU-MDR.
Section “5.1 product changes” is only applicable for section 4.4 of Annex II of the Directive 93/42/EEC (which is not applicable for Class IIa and IIb devices, see section 7 of Annex II) and section 6 annex III (which cover EC type-examination certificate). Therefore, it appears that the list of changes to product are only applicable in the cases where there is a specific product related certificates provided.
Sections “5.2 Quality System Changes” and “5.3 Changes to the product range” these refer to the relevant sections of the old MMD that discuss changes to the QMS and the product ranges, this is in line with the current EU-MDR Annex IX path way.
Therefore at least one of the major notified bodies indicates that although the manufacturer needs to report all changes they only have to consider the criteria in section 5.2 and 5.3 to asses if a change is deemed substantial as considered in section 2.4 of annex IX. We have successfully support our client receive notified body approval to make changes to the indications for use and to the core software algorithm without prior technical file review.