Is the FDA review much faster than EU CE‑marking? Not really — or it depends…

One of the key challenges for MedTech startups is the long duration of the regulatory clearance process. You have developed your product, completed the bench and safety testing, collected the clinical evidence, and are ready to sell… but now you must wait for someone to review your file before you can bring the product to market. During this time, you are burning runway while essentially twiddling your thumbs. So, the strategic question many startups face is: where should we seek clearance first?

The common belief is that if you want fast market clearance for a medical device, you are better off going to the US. Many articles (and your favorite chat bot) will tell you that clearance in the US takes around 3 to 6 months for medium‑risk devices (510(k)) and up to a year for high‑risk devices (PMA). By contrast, the often‑quoted CE‑marking timelines are 9 to 24 months or more. So, the conclusion seems obvious: the FDA route is faster! But on the other hand, we recently helped a startup achieve CE‑marking for its first product in 4.5 monthsafter submission (including QMS certification).

So, what do the numbers actually say? In this article, we share our original analysis of FDA clearance timelines, with a particular focus on the timelines for companies launching their first product. We also discuss the limitations and biases in the readily available data.

On the EU side, we attempted to trace the sources of the quoted timelines and found that the EU and US timelines measure different things, making it hard to compare them directly. We also explore some of the reasons behind longer EU timelines, which are at least partly driven by how notified bodies structure their review processes. Next week’s follow-up article will explore in more detail the elements that will impact the optimal choice for several specific cases.

Analysis of the FDA review timelines

In general, for medium‑ to high‑risk devices there are three different pathways to receive clearance. Medium‑risk (Class II) devices can follow two processes. The 510(k) process is for devices where there is already a legally marketed predicate device with the same intended use and no technical differences that raise new questions related to safety or performance. If a new device is deemed substantially equivalent, the FDA presumes a positive risk‑benefit profile and typically completes its review in 90 days. For significant changes to a device, companies must resubmit a 510(k), but those reviews are usually faster since there is less debate around the predicate and required evidence.

For medium‑risk Class II devices without a predicate, the De Novo process is required. Because there is no predicate, the company must demonstrate to the FDA that the novel device has a positive risk‑benefit profile compared to the standard of care. This involves a more extensive review, and the FDA aims to complete it in 150 days.

For high‑risk Class III devices, the Pre‑Market Approval (PMA) process is used. Initial submissions take at least 180 daysand can take longer if more detailed clinical reviews are required. Any changes to a Class III device also require PMA approval, though these supplemental PMAs are typically reviewed more quickly.

In all cases, the review times mentioned represent the net FDA review time—the time the FDA has the file. When they identify deficiencies, they issue a formal request for additional information. This stops the review clock, and manufacturers have up to 180 days to respond.

For all approved submissions, the FDA publishes a summary that includes the original submission and clearance dates. At MD squared, we have built a detailed database that allows us to analyze this publicly available information in far greater detail than what is generally accessible online. This provides novel insight into actual review times in practice, especially for first‑time submitters.

Below, we analyzed all approved submissions from 2015 to 2025 and visualized the distribution of review times for 510(k), De Novo, and PMA processes. For each pathway, we also show timelines specifically for first‑time submitters.

Artikelcontent

 
Original analysis – Copyright MD squared

Box plot of the time (in days) between submission and approval for the different regulatory pathways. The center line indicates the median; the box shows where 50% of values fall. Times for first‑time submitters are shown separately. Only initial PMA submissions are considered; supplemental PMAs are excluded.

From this analysis, it is clear that there is a significant spread in review durations. For first‑time submitters, timelines are slightly longer (8–40 days more) than for companies that have submitted before, though the effect is smaller than we expected.

The commonly quoted 3–6-month range for 510(k) approvals is reasonable. For first‑time submitters, half of approvals occur between about 3 and nearly 9 months.

However, for novel medium‑risk devices requiring the De Novo pathway, review times are substantially longer. Half of De Novo reviews for first‑time submitters take between 7 and 14 months.

For high‑risk Class III devices, the median review time is around a year, meaning the often‑mentioned “up to a year” estimate is somewhat optimistic, as clearance takes longer than a year in many cases.

Survivor biases

A critical caveat is that these values only consider submissions that were ultimately approved. If the FDA issues questions that cannot be answered within the 180‑day clock stop period, the submission is considered withdrawn. Any later re-submission is treated as a new submission. This creates strong survivor bias: average timelines for “approved” files may significantly underestimate the actual time companies spend in the process.

To estimate abandoned submissions, we use the fact that each submission type has a sequential case number that restarts each year (e.g., K230001 is the first 510(k) submission of 2023). For each year, we identify the highest observed case number to estimate the total number of submissions. Comparing this with the number of approved files from that year gives us an estimated approval rate per pathway.

Artikelcontent

Original analysis – Copyright MD squared

Estimated percentage of approved 510(k) (blue), De Novo (red), and PMA (green) submissions from 2015 to 2025. Note: values for 2024 and 2025 are underestimated because some submissions are still under review.

The resulting graph shows that roughly 80% of 510(k) submissions are approved. For PMAs, this varies more (due to lower submission volume), but on average only about 65% are approved.

De Novo submissions fare even worse: fewer than half of the submissions each year are approved. This puts the average approval times in a different perspective. Particularly for medium‑risk devices without a predicate, the De Novo process is not only long but also far from guaranteed to succeed.

EU numbers?

So, what about the EU? There is no detailed publicly available dataset comparable to the FDA’s. Online, various publications quote timelines such as:

  • “Typically, it can take between 9 months to 24 months for conformity assessment to be completed, but in some cases may be longer.” Source: Team NB, the organization representing Notified Bodies – LINK
  • “Typically, the process takes between 12 to 18 months to complete.” Source: DEKRA, a large notified body – LINK

But it is unclear what these numbers are based on or what they actually measure. The EU regularly surveys notified bodies and publishes a detailed presentation and even a dashboard.

Artikelcontent

Source: Updated Notified Bodies Survey Certifications and Applications 2025-08-01

The most relevant information regarding time to first certification appears on slide 30. It shows that 60% of notified bodies report timelines of 13–18 months. But there are several caveats:

  • The footnote clarifies that this is the time from signing the written agreementto issuance, not from submission of the technical file. In recent years, when capacity was limited, companies often signed months (or even years) before being ready to submit.
  • Each notified body appears to provide a singlevalue, not a distribution across individual devices.
  • Notified bodies report that most initial submissions are incomplete. Only 15 out of 50 notified bodies state that more than half of the files they receive are complete. (see: slide 29 of the report)

What does this mean for start‑ups submitting in the EU?

To understand these numbers, it is important to know how notified bodies operate. They have no mandatory or target initial review times, and until recently, capacity was heavily constrained. When a company submitted a technical file, the start date of the review was uncertain. And since a company can only submit to onenotified body, timing was largely outside their control.

Most (all?) notified bodies offer dedicated review lanes where they commit to starting the review on a specified date and respond within specific time frames after each round of questions. But this requires the manufacturer to commit to being ready by that date. These VIP lanes come at a significantly higher cost, but for startups with limited runway, paying extra to ensure the review actually begins can be worth it.

However, missing the slot could mean being charged anyway. Rescheduling often requires confirmation up to 60 days in advance. This created pressure, especially for companies still completing (clinical) testing, to submit an incomplete file just to keep their slot. They hoped to supply missing data after the first review round. Furthermore, since the deadline for responding by manufactures on question raised by the notified bodies are less stringent, increasing the total review timeline. This naturally leads to incomplete files, longer reviews, and frustration on the part of notified bodies and companies a like.

Today, capacity at notified bodies appears to be increasing again, reducing the need for early slot‑locking. To minimize review time, it is crucial to ensure the technical file is high‑quality, consistent, and complete. Under these conditions, and with notified‑body support, CE approval in 4.5 months from the moment of submission is indeed possible, as we recently demonstrated for a client.

Conclusion

Based on our analysis of US clearance data and a clearer understanding of EU timelines, the story is more nuanced than “FDA fast, EU‑MDR slow.”

In the US, timelines and effort vary significantly depending on the chosen pathway. For a medium‑risk Class II device requiring the De Novo process, the road is challenging: over half of submissions appear to be abandoned, and successful ones often take a year to clear. This makes it critical to align with the FDA on appropriate pathway and required evidence before submission. Fortunately, the FDA offers the Q-submission process that gives manufactures to discuss these topics in an early stage.

In the EU, far less detailed information is available, and commonly cited timelines start at the moment a contract is signed, sometimes long before a file is ready. With increasing capacity at notified bodies, and with a well‑prepared, complete technical file, fast EU clearances are indeed achievable.

So, which is faster—FDA or MDR? It depends. Not only on the regulatory pathway, but also on the time invested before submission. FDA and EU‑MDR often differ in the evidence they require, and these differences can profoundly shape the overall path to market. Careful analysis is essential to allow informed choices to be made on this critical topics.