If A Regulatory Model Works for AI in Planes, Trains, and Now Machines—Why Not for Medical Devices?

Over the past few days, I have tried to sit with the decision by the European powers‑that‑be to leave the status of medical devices unchanged under the AI Act. I’ve discussed it with people both inside the medical device world and with others who look at this from the outside. What keeps surfacing—both in those conversations and in my own thinking—is a sense of puzzlement and a growing frustration that is difficult to shake.

This article is an attempt to make sense of that feeling. I do not know who will read this, or where it might land. But I do know that the implications of this decision are too important to dismiss with a casual confidence that “the impact will probably be limited” or that “we can always fix it later.” In many regulatory domains, that assumption may even be reasonable. In medical devices, I am far less convinced.

I am writing partly to better understand how we arrived here, partly to explain to outsiders why many medical device experts are concerned, and partly to test whether this concern can be articulated clearly enough to travel; to be shared, discussed, and, ideally, considered by those who shape policy. If nothing else, it helps me clarify how we might move forward from this point.

And not all hope is lost! There is a parallel and ongoing discussion around the proposed targeted update of the EU‑MDR and IVDR, which includes a similar proposal aimed at reducing overlap between those regulations and the AI Act. That discussion may yet prove decisive. But to engage with it meaningfully, we first need to be honest about what is at stake.

We begin with a brief overview of how medical devices are regulated under the EU‑MDR, highlighting that many of the core concepts introduced by the AI Act (such as quality management, risk management, transparency, and post‑market surveillance) are already deeply embedded in the existing framework. This naturally raises the question: if medical devices already meet these expectations, what exactly is added by subjecting them fully to the AI Act as well?

We then look at how AI is actually used in medical devices today. This is important, because while medical devices undeniably touch on health (a fundamental right) not every AI application in this domain carries the same types or levels of risk that originally motivated the high‑risk categorization in the AI Act. AI‑driven medical systems already fall under the EU‑MDR and are subject to significant oversight even without the AI Act.

Finally, we examine what it really means to move medical devices from Annex I Section A to Section B of the AI Act. This is not about exempting medical devices from AI‑specific requirements, but about understanding how other highly regulated sectors are treated, and why a different integration approach is considered acceptable there. Placed in that context, the current position on medical devices starts to look less self‑evident and this clearly deserves closer scrutiny.

Regulatory oversight on medical devices in the EU

Any device or software intended to screen, triage, diagnose, or treat a person (see note 1) falls under the Medical Device Regulation (MDR, Regulation (EU) 2017/745; see note 2 for IVDR). The level of regulatory oversight depends on the potential risk posed by the device.

Most—if not all—AI‑based medical devices “provide information which is used to take decisions with diagnosis or therapeutic purposes” (see note 3). As a result, they are generally classified as medium‑risk (Class IIa or IIb) to high‑risk (Class III) devices.

All medical device manufacturers must implement a Quality Management System (QMS), typically in accordance with ISO 13485, which is harmonized with the EU‑MDR. This requires a controlled development process and the implementation of an effective risk management framework, usually based on the harmonized ISO 14971 standard.

ISO 14971 mandates a structured process to identify, assess, and minimize risks across the entire product lifecycle—from initial development through production and post‑market activities. All identified risks must be reduced as far as possible. For software‑based devices, cybersecurity has become a critical component, with dedicated standards and guidance reflecting the type of data processed and the environments in which devices are deployed.

Device performance must be verified through internal technical testing to demonstrate that design requirements are met and risk controls are effective. In addition, every medical device must undergo a clinical evaluation to assess its clinical benefit, safety, and performance based on scientific literature and clinical investigations.

To generate performance data, most devices require one or more clinical investigations conducted in real clinical settings. These studies support an overall assessment of safety and performance and the acceptability of the benefit‑risk ratio compared to alternative treatments.

Manufacturers must also define a Post‑Market Clinical Follow‑Up (PMCF) plan to monitor device performance once it is on the market. Depending on the risk profile, this may range from complaint monitoring to additional post‑market clinical studies. All evidence is compiled into a comprehensive technical file, often comprising thousands of pages per device.

To place a Class IIa or higher‑risk device on the market, manufacturers must involve an external notified body. This includes a multi‑day on‑site audit of the quality management system and an in‑depth review of representative technical files. Reviewers assess documented evidence against EU‑MDR requirements and applicable harmonized standards, raising findings where information is missing or insufficient.

Certification typically takes between 6 and 18 months and costs EUR 50,000 to EUR 125,000 in notified‑body fees alone.

 

After market entry, notified bodies maintain continuous oversight through annual audits, reviews of technical file updates, and unannounced audits at least once every five years. Manufacturers pay tens of thousands of euros annually to maintain certification.

With the introduction of AI and machine learning, the medical device ecosystem—manufacturers, notified bodies, and standards organizations—has adapted by developing additional guidance and requirements within the existing MDR framework. Examples include the ongoing development of the technical report “24971‑2.2 Medical devices — Guidance on the application of ISO 14971 Part 2: Machine learning in artificial intelligence” and MDCG guidance 2025‑6 on the interplay between MDR, IVDR, and the AI Act.

In addition, Team‑NB has published detailed questions that notified bodies may raise when reviewing AI‑based medical devices, grounded in existing EU‑MDR obligations. Because the MDR requires manufacturers to implement state‑of‑the‑art methods, notified bodies can already require adoption of new standards and guidance as they emerge.

The system is not perfect, but this continuous and structured oversight enables notified bodies to monitor and challenge how new technologies are implemented, supporting the ongoing safety and performance of AI‑based medical devices even as regulatory expectations evolve.

Regulatory oversight under the AI Act

The stated aim of the EU AI Act is to create a single, EU‑wide legal framework that ensures artificial intelligence is safe, transparent, ethical, and respectful of fundamental rights, while still enabling innovation and competitiveness.

The AI Act addresses the uneven and sometimes serious risks posed by AI systems, including discrimination, lack of transparency, unsafe autonomous behavior, and excessive surveillance. Low‑risk AI systems—such as spam filters or AI used to improve photo quality—are largely left unregulated. High‑risk AI systems, such as algorithms used for CV screening or creditworthiness assessments, face strict obligations due to their impact on rights and safety. Certain AI practices, including real‑time biometric surveillance in public spaces and government‑run social scoring systems, are prohibited outright.

For high‑risk AI systems, Chapter III of the AI Act defines requirements for both the systems themselves and the organizations that provide or deploy them. Key requirements relate to quality management systems, risk management, data and data governance, human oversight, transparency, cybersecurity, and post‑market monitoring.

The AI Act deliberately defines these requirements at a high level, as they must apply across many types of AI systems. To operationalize them, the European Commission issued a standardization request to European standards organizations to develop harmonized standards translating these principles into concrete and implemantable requirements.

These standards aim to provide practical guidance while maintaining flexibility across sectors. Once cited in the Official Journal, compliance with these standards provides presumption of conformity with the Act.

 

This process is still ongoing. The original AI Act articles (9–15, 17, and 72) contain roughly 4,000 words and 67 “shall statements,” which are now being expanded into more than 10 entirely new standards.

As a member of the Dutch standardization committee on Artificial Intelligence and Big Data, I have actively followed this drafting process. It has been a massive undertaking, initially delayed but now delivering its first standards—covering QMS, risk management, and cybersecurity—for public consultation.

Each of these draft standards spans more than 50 pages and contains between 70 and 130 “shall statements.” It is already clear that several of these requirements go beyond what is explicitly stated in the AI Act itself. Implementation and documentation of this implementation of all these new standards will be a mammoth task for each manufacturer.

High‑risk AI systems must undergo conformity assessment by a notified body (Article 43 and Annex VII) before being placed on the market. Manufacturers must document in detail how they meet requirements across all applicable standards, and notified bodies must review this evidence. Only after a successful review may the manufacturer issue an EU Declaration of Conformity (Article 47) and affix the CE mark.

Since none of the standards are finalized and no notified bodies have yet been appointed, timelines and costs remain unclear. However, experience with the EU‑MDR suggests it will take time for notified bodies to be appointed, build capacity, and for shared interpretations to emerge.

It is important to recognize that under many EU regulatory frameworks, manufacturers are allowed to self‑certify compliance and bring products to market, with any gaps, ambiguities, or unintended consequences addressed over time through market surveillance and enforcement discretion. This approach allows regulation and practice to mature in parallel, without immediately obstructing innovation or access to new technologies.

The situation is fundamentally different for high‑risk AI systems. Because these systems require mandatory third‑party conformity assessment before they can be placed on the market, any uncertainty, immaturity, or lack of practical consensus in the law or the supporting standards does not merely create friction, it will act as a hard barrier. Products simply cannot enter the market until those uncertainties are resolved.

This distinction matters enormously for medical devices, since they are already continuously assed by notified bodies. What might be framed as an imperfection that can be corrected later in other policy areas can, in this context, translate directly into innovation being stalled or abandoned. That outcome may well be consistent with the strict‑by‑design philosophy of the AI Act. But if it applies equally and mechanically to medical devices, it risks having an outsized and potentially unintended impact on a sector where delay itself can carry real consequences.

Impact of the AI Act on medical devices

Under the AI Act, medical devices are listed in Annex I Section A, meaning all AI‑enabled medical devices are classified as high‑risk AI systems. To understand the impact, it is important to look at typical AI use cases in medical devices. Broadly, these can be grouped into three categories:

Image or data processing algorithms.These process medical images (pathology slides, X‑ray, MRI, ultrasound) or patient data streams (ECG, EEG, lab results) to identify or classify diseases or provide treatment support. Historically, these tasks used classical algorithms. Also with AI-systems, physicians still control clinical decisions and workflow. Existing medical device regulations are well‑suited to assess the benefits and risks of these systems, including training data appropriateness, input compatibility, and potential biases.

AI systems supporting triage or diagnostic pathway selection.These systems assess patient parameters to provide risk classifications and recommendations. Classical scoring systems already exist (e.g. Framingham Risk Score, NEWS score). Black‑box AI systems may perform differently across populations, potentially affecting fundamental rights, and thus require additional focus during risk management and clinical evaluation.

Diagnostic expert systems. These newer systems use unstructured input (e.g. symptoms described in text) to generate differential diagnoses and are often built on LLMs combined with curated medical data. These are already classified as medical devices under the MDR. “Prof. Valmed” is a Class IIb certified device that supports diagnosis, treatment, or monitoring decisions. Competing products (ChatGPT for Physicians and OpenEvidence) that are being offered on the (unregulated?) US market are already being blocked from entering the EU under the EU-MDR.

 

Regardless of application or risk class, classification as high‑risk under the AI Act means manufacturers must implement all AI Act requirements. Many—such as QMS, risk management, cybersecurity, and post‑market monitoring—closely align with MDR requirements. Others, including transparency, are already addressed through labeling and instructions for use.

The AI Act also introduces explicit data governance requirements. While not named as such in the MDR, patient data collection already requires clinical investigations, ethics approval, and informed consent. Human oversight requirements likewise need appropriate consideration, depending on the specific use case.

Article 43.3 of the AI Act specifies that medical devices must follow both the MDR conformity assessment andassessment of compliance with all AI Act product requirements.

So, if there is so much overlap, what is the problem?

One could argue that since medical device manufacturers already meet most AI Act requirements, they should simply comply. The challenge is that this requires implementing and proving compliance with more than 10 new AI Act standards—many overlapping with existing medical device standards.

These standards are still immature. Experience from the medical device domain shows it typically takes 5+ years and at least one major revision for standards to become practical and workable.

A further challenge is that the medical device industry is uniquely fragmented. Over 37,000 medical technology companies operate in Europe, and around 90% are SMEs, most with fewer than 50 employees (source). These companies lack resources to track and influence standardization. The resulting regulatory uncertainty and increased burden are already harming investment in AI‑driven medical startups, with long‑term consequences for EU innovation.

Planes, Trains, and Automobiles (and now machines)

Interestingly, several other heavily regulated sectors are listed in Annex I Section B rather than Section A, including (unmanned) planes (2018/1139), trains (2016/797), and automobiles (2019/2144). In the latest AI Omnibus update, the Machinery Regulation (2006/42/EC) was also moved from Section A to Section B.

For these products, sectoral legislation is amended to incorporate relevant AI Act requirements. They do notrequire a separate AI Act conformity assessment; requirements are addressed within their existing certification frameworks.

This raises a fundamental question: why is this approach acceptable for planes, trains, and self‑driving cars—but not for medical devices?

Yes, medical devices impact health, a fundamental right. But they are already subject to some of the strictest regulatory oversight, including third‑party conformity assessment and continuous surveillance. Implementing any remaining gaps with the AI-act can be implemented via an update of the EU-MDR or aditional sector specific standards and guidance. What does this additional independent legislative layer achieve that cannot be addressed through targeted MDR updates?

This question lies at the heart of my (and many medical device experts’) frustration and disbelief.

So how to proceed?

Not all hope is lost—but timing and clarity matter.

On December 16, 2025, the European Commission proposed a targeted revision of the EU MDR and IVDR aimed at simplifying rules, reducing administrative burden, and strengthening innovation capacity. As argued in a previous article, this proposal again includes moving medical devices to Annex I Section B of the AI Act and integrating the relevant AI requirements directly into the MDR framework.

In the EU decision‑making process, three institutions must align for such a change to materialize: the European Commission, the European Parliament, and the European Council. The Commission has already taken a clear step toward simplification. The European Parliament’s final position is still pending, but its proposed update to the AI Act indicates support for moving the MDR to Annex I Section B, alongside other sectoral legislation.

 

This leaves the position of Member State governments as the decisive factor. At present, it remains unclear which governments favor keeping medical devices in Annex I Section A, and on what concrete grounds. That lack of transparency makes it even more important that the discussion moves beyond general statments and toward a substantive assessment of proportionality, feasibility, and long‑term impact.

What is needed now is not a debate framed in absolutes—for or against regulation—but a more precise conversation about how regulation interacts with existing frameworks that are already doing much of the intended work. For medical devices, the question is not whether AI requires oversight, but whether duplicative conformity assessment is the right mechanism in a sector that already operates under continuous third‑party surveillance.

For manufacturers, developers, and innovators, this creates a period of real uncertainty. Decisions made now, about investment, product roadmaps, clinical strategy, and regulatory pathways, may lock in consequences long before legal clarity emerges. Navigating this moment requires a detailed understanding of both regulatory regimes, the evolving standardization landscape, and how notified bodies are likely to interpret overlaps in practice.

At MD squared B.V., this is precisely where we have been operating for years: at the intersection of medical device regulation, AI‑based technologies, and practical implementation. We work with teams that are trying to make responsible decisions under imperfect conditions; helping them interpret regulatory signals, prepare for multiple plausible futures, and remain compliant without losing momentum.

More broadly, I hope this article contributes to a more grounded and open exchange between regulators, policymakers, and the medical technology community. The choices being made now are not merely technical. They shape who can innovate, at what pace, and under what conditions—and those effects will be felt long before any future “fix” can be applied. Getting this right is not about lowering standards; it is about ensuring that they are applied where they add value, rather than friction, in a field where delay itself is not a neutral outcome.

Notes

Note 1:This definition paraphrases Article 2.1 of the EU‑MDR.

Note 2: Most AI‑based products will fall under the Medical Device Regulation (MDR). However, certain devices may instead qualify as in vitro diagnostic medical devicesif they act on, or provide information based on, specimens that have been taken from the human body. Examples of such specimens include urine, blood, or biopsy samples. These devices are regulated under the IVDR (Regulation (EU) 2017/746). The concepts discussed in this article are also applicable to devices regulated under the IVDR; however, some of the specific regulatory clauses referenced will differ from those under the MDR.

Note 3: See “MDCG 2019‑11 Rev.1 Guidance on Qualification and Classification of Software in Regulation (EU) 2017/745 – MDR and Regulation (EU) 2017/746 – IVDR” for detailed guidance on determining whether a software device falls under the scope of the EU‑MDR and how it should be classified. As stated in the guidance, “The wording ‘intended to provide information which is used to take decisions with diagnosis or therapeutic purposes’ describes, in very general terms, the ‘mode of action’ which is characteristic of all MDSW. Therefore, this sub‑rule is generally applicable to all MDSW,” and such devices should therefore be classified as at least Class IIa. The guidance further explains that “[devices intended] to prevent the risk of illnesses or pathologies by analysing physiological parameters [..] can be considered as a device providing information which is used to take decisions with diagnosis purpose [..].”This means that devices intended for initial screening or for use in a preventative setting would also be classified as at least Class IIa.

Erik 02

Erik van Dijk

Managing Partner