In recent weeks, we have supported several clients in developing or refining their Quality Management Systems (QMS). These companies were at very different stages of their company lifecycle, and it was striking to see where their QMS added value, and also where it did not. The experiences highlight an important lesson for MedTech companies: the value of a QMS depends heavily on the stage you are in and the way you scope and implement it. Many companies get this balance wrong, which reduces the practical value of their QMS and drains energy from their teams.

Below are three premises that are often overlooked:

  1. Detailed procedures only add value for processes you perform regularly and consistently.
  2. You can only write an efficient, realistic procedure after running the process a few times.
  3. Some procedures exist purely to satisfy ISO 13485 and provide little business value to a startup.

Taken together, these lead to a conclusion that can feel counterintuitive: For early-stage companies developing their first product, it is often best to implement only the minimum procedures required for compliance and rely on experienced people and solid templates to create high-quality records. For more mature organizations, the priority shifts to defining processes that actually work for the team, supported by the tools they use daily, and only then incorporating the regulatory requirements.

Below are three recent cases illustrating how different approaches create or limit value.

Case 1: Startup Completing Its First Product

We recently worked with a startup preparing the technical file for its first product. The team had invested considerable effort in building its QMS and had passed Stage I of their ISO 13485 audit more than a year earlier. However, they could not pass Stage II because the Notified Body could not assess the effectiveness of their QMS—simply because development was not yet complete. (see previous post)

Despite having all development procedures documented, the team struggled to produce documentation at the required level of quality. Their challenges were highly practical:  How detailed should requirements be?  How should they classify risk severity?  How should they define “serious” versus “minor”?  How do you create the “indicative list and specification of parameters to assess benefit-risk” required in the clinical evaluation plan?

To address these questions, we provided hands-on guidance and detailed record templates. And we largely ignored their existing QMS procedures. Not because the procedures were wrong, but because theoretical, generic procedures do not help a team going through these tasks for the first time.

Once the technical file was submitted, we revisited their procedures to clean them up. For example, their QMS required a dedicated Infrastructure Plan, and an outdated draft existed, but the team at that time did not yet know what infrastructure their product actually required. Because this was their first (and for now only) product, we had embedded the relevant ISO 13485 Clause 6.3 requirements directly into their design transfer plan and report when building the technical file. This ensures that it had the most up to date data, we could add rationales why this was sufficient for their product. No needless doubling of information that would need to be kept up to date.

This was a clear example of an early-stage team needing fewer procedures, better templates, and practical support—not a detailed design and development process they were not yet equipped to use.

Case 2: Scale-Up Releasing Its Second and Third Software Device Versions

Last year, we assisted another company in submitting the technical file for the first release of their Software as a Medical Device. The company had invested heavily in developing a working algorithm and a clinically usable first version, leaving limited bandwidth for process development.

Working with the QA manager, we documented a lean but compliant process, focusing on ensuring the records met all ISO 13485, ISO 14971, and EU-MDR requirements. We held formal design input reviews to confirm requirements were complete, non-contradictory, and testable, and that risk control measures were properly captured. The procedures did not prescribe exactly how or when these reviews should happen or who should lead them—this would have been premature.

Because the records were strong, the company had no issues certifying both their QMS and product.

Now that the team has completed its second and third releases, they finally understand what works for them. They are defining the operational details: who does what, when key reviews should occur, and how to structure documentation. These procedures now add real value by making releases smoother and onboarding easier.

Had we tried to define detailed procedures before the team had any real experience, the result would have been theoretical, unhelpful and distracted from making sure the product actually works.

Case 3: Contract Development Company Expanding from ISO 9001 to ISO 13485

We are also working with a contract development firm that already holds ISO 9001 certification. Because many of their clients are in MedTech, they want to upgrade to ISO 13485. At the same time, they are not satisfied with their current working processes.

Before touching their QMS, we discussed how they actually perform their work: which tools they use for product data management, how design reviews (including CAD) are conducted, and how they track changes. On top of these real processes, we will add the ISO 13485 Clause 7 product-realization requirements.

For other clauses, we will keep things lean. For example, management review is a requirement, but given the company’s small size, the formal procedure as required under ISO13485 will add little business value. They still need it for compliance, but it should be kept minimal.

Conclusion

A well-structured QMS can add real business value far beyond simply holding an ISO 13485 certificate. But this only happens when timing and scope are handled correctly. A QMS introduced too early, or designed in too much detail too soon, can slow down an organization and encourage a checklist mentality that does nothing to improve product quality or patient safety.

A right-sized QMS, introduced at the right time, can instead empower teams, improve consistency, and support safe and efficient product development.

To comment and share your thoughts join the conversation at LinkedIn!

Related posts:

Public Gatekeepers, Private Returns

The commercial incentives shaping EU medical device conformity assessment Key Messages The EU’s decentralized conformity assessment model relies on independent, mostly commercial notified bodies to

Read More »

Don’t hesitate to ask us your questions.

Drop an e-mail or contact us via LinkedIn

This website uses cookies to ensure you get the best experience on our website. If you want to get more information about this, please view our privacy policy page.