FDA PCCPs: Do they really help you move faster, or just add complexity?

If you’re developing a software medical device (particularly one incorporating AI/ML) you’ve likely heard that Predefined Change Control Plans (PCCPs) are the answer to one of the biggest frustrations in US regulation: the need to submit a new 510(k) for every meaningful update.

The narrative is compelling. PCCPs are often described as a way to “future-proof” your device, allowing you to iterate faster, avoid repeated submissions, and bring the speed of software innovation closer to regulatory reality. For many, they promise a more efficient path through the FDA.

But how much of that promise holds up in practice?

At their core, PCCPs allow manufacturers to define and validate certain types of changes upfront, so they can be implemented later without requiring a new submission. In principle, this creates room for faster iteration while staying within regulatory guardrails.

In reality, the picture is more complex.

Including a PCCP introduces additional upfront work and strategic trade-offs. Does it complicate your initial submission? Is it worth including from the start, or better added in a follow-on 510(k) once you have a cleared product? And ultimately—does it actually reduce the need for future submissions?

These are not just regulatory questions; they directly affect how you sequence development, manage runway, and make decisions about when—and how—to update your product.

In this article, we analyze the complete set of 114 Class II devices cleared with a PCCP to date (as of May 25, 2026) to understand how this mechanism is being used in practice. We examine which types of devices include PCCPs, what kinds of changes they cover, and how frequently they are applied across product categories.

Based on this analysis, we separate perception from practice. We outline when a PCCP can genuinely add value and when it is unlikely to deliver on its promise. We also highlight alternative approaches that can help streamline regulatory pathways without adding unnecessary complexity or constraining future product development.

PCCP introduction

Unlike traditional hardware devices, software-based medical devices, especially those incorporating AI/ML, are built to iterate rapidly and continuously evolve after deployment. Updates are more frequent, incremental, and often essential to maintaining or improving performance. However, the US regulatory framework was largely built around hardware products, where changes occur less frequently and are easier to bound.

Under existing FDA guidance, many software modifications require a new 510(k) submission. This includes changes intended to improve performance, such as retraining an algorithm or to expand compatibility with new inputs, devices, or use environments. While this approach ensures continued oversight, it can also create significant friction for products that are designed to evolve over time.

Recognizing this tension, the FDA introduced an alternative mechanism: Predefined Change Control Plans (PCCPs). A PCCP allows manufacturers to prospectively define the types of changes they intend to make, along with the associated verification and validation activities needed to demonstrate that the device remains safe and effective. The PCCP is submitted and reviewed as part of a marketing application (510(k), De Novo, or PMA), enabling certain future changes to be implemented without requiring a new submission.

Although PCCPs are often discussed in the context of AI/ML-based devices, they are not limited to this category. The FDA has published guidance specifically addressing AI/ML-enabled devices, as well as broader (draft) guidance extending the PCCP concept to other types of medical devices.

The concept of PCCPs has evolved over several years. The initial proposal dates back to 2019, with the first device including a PCCP (DEN190040) cleared on February 7, 2020. In this article, we analyze that device together with 113 subsequent Class II clearances that incorporate a PCCP.

As of May 25, 2026, a total of 114 devices have been cleared with a PCCP: 108 through the 510(k) pathway and 6 via the De Novo program. The figure below illustrates the slow adoption of PCCPs in the early years, followed by a more recent increase in usage.

Adoption remained limited in the early years. However, starting in 2025, the number of devices incorporating PCCPs increased consistently. This trend aligns with the publication of draft PCCP guidance documents in 2024, which appear to have provided greater clarity and confidence for manufacturers to adopt this pathway.

The figure below shows that PCCPs are beingadopted across nearly all FDA medical specialties and review panels. To date, 13 of the 19 review panels have cleared devices with a PCCP; those that have not yet done so include Dental, General Hospital, Hematology & Ophthalmic, Physical Medicine, and Toxicology.

Radiology is by far the most active area, followed by Cardiology and Microbiology. The strong presence of radiology and cardiology is expected, as these areas include many software-driven devices—particularly imaging and diagnostic tools—where performance improvements depend on frequent algorithm updates.

The relatively high number of PCCPs in microbiology is driven by a different use case. In many in vitro diagnostic (IVD) devices, PCCPs are used to address future revisions to device labeling in response to breakpoint updates recognized on the FDA’s STIC webpage. This illustrates that PCCPs are not limited to enabling iterative software improvements, but can also support more predictable, standardized updates in non-AI/ML contexts.

An AI-based review of product descriptions shows that the majority of devices (62%) include an AI/ML component. A further 20% are in vitro diagnostic (IVD) devices. Notably, two devices fall into both categories, as they use AI/ML to analyze pathology images and are therefore counted in both groups.

 

The remaining 21 devices (18%) fall into an “other” category. This group is highly heterogeneous and includes products such as surgical gowns (K252301), an endoscope (K251562), and bone screws (K243943).

This distribution highlights that PCCPs are already being applied well beyond AI/ML-driven devices.While much of the initial discussion around PCCPs focused on enabling iterative learning systems, their broader adoption across diverse device types is aligned with the FDA’s intent to create a flexible framework for managing predictable post-market changes more generally.

A further AI-based analysis of PCCP summaries examines the types of changes manufacturers anticipated at the time of submission.

The results show that most PCCPs focus on modifications to AI/ML algorithms—primarily updates to the models themselves and retraining with new data. The next most common category involves expanding compatibility, such as adding new input sources or supported devices. 

 

It is important to note that a single PCCP can cover multiple types of changes, meaning these categories are not mutually exclusive but reflect the range of planned modifications within each device.

How popular are PCCPs?

As shown in the first figure, PCCPs are still a relatively recent development, and the current dataset (114 clearances) is small compared to the broader regulatory landscape. Each year, more than 3,000 Class II devices are cleared through the 510(k) and De Novo pathways, putting the adoption of PCCPs into perspective.

A more relevant question, therefore, is how frequently PCCPs are used within specific product categories. To explore this, we analyzed devices cleared after January 1, 2025 (82 in total), focusing on product codes where at least two devices included a PCCP. The figure below shows, for each of these product codes, the percentage of clearances that incorporated a PCCP.

This analysis shows that, in most product codes, the majority of devices are still cleared without a PCCP. There are, however, a few notable exceptions.

  • QWO(AI-based radiology analysis) is a relatively new product code where a single manufacturer has cleared multiple devices with a PCCP, building on its own initial De Novo classification.
  • QJU(also AI-based radiology analysis) includes the first device ever cleared with a PCCP (Caption Guidance). Because the original De Novo included a PCCP, most subsequent devices in this category have followed the same approach.
  • QJI(automated glycemic controllers) shows broader adoption across multiple manufacturers, indicating that PCCPs can gain traction when a clear regulatory precedent is established.

It is important to note that this analysis is subject to selection bias: product codes with only one (or no) PCCP clearance are not included. Even with this limitation, the overall picture is clear—PCCPs are being used consistently in only a small subset of eligible submissions.

Whether this reflects cautious adoption, limited applicability, or remaining uncertainty will likely become clearer over the coming years as both guidance and experience continue to evolve.

When do companies add a PCCP and does it slow you down?

For most Class II devices in the US, regulatory strategy is heavily shaped by the ability to identify an appropriate predicate and demonstrate substantial equivalence. In a typical 510(k), this means showing that your device is sufficiently similar to an already cleared product in terms of intended use, technological characteristics, and performance.

For first-of-a-kind or highly innovative products, this can be challenging. Suitable predicates may not exist, or there may be significant technical differences that require additional justification and supporting evidence. As a result, initial submissions are often more complex, time-consuming, and uncertain compared to follow-on submissions that can rely on a company’s own previously cleared device as a predicate.

This dynamic raises an important strategic question: at what point in the product lifecycle do companies choose to introduce a PCCP? Is it something that is typically included in the initial submission, or added later once a regulatory baseline has been established?

To explore this, we analyzed the predicates used in PCCP-containing clearances and assessed whether at least one predicate device originated from the same manufacturer. The results are shown in the figure below:

This distribution suggests that, in practice, PCCPs are most often introduced in follow-on submissions rather than in initial clearances.Once a company has an FDA-cleared device, it becomes significantly easier to define the scope of future modifications, build on existing validation approaches, and align with prior regulatory expectations. Incorporating a PCCP at this stage can therefore be more straightforward and lower risk.

This naturally raises the question of whether this pattern is driven by the “cost” of including a PCCP. Intuitively, adding a PCCP to an initial submission (where substantial equivalence is already more difficult to establish) could increase complexity and potentially slow downthe review process. In contrast, adding a PCCP in a subsequent 510(k), where the predicate is well understood and internal data is available, may be expected to have less impact.

 

To test this hypothesis, we analyzed clearance times within the QIH product code. This group includes advanced radiological image processing software, such as AI-based tools, and was selected because it contains the largest number of devices with PCCPs (19 in total: 14 using a predicate from the same company and 5 using a predicate from a different company).

The figure below compares clearance timelines across two dimensions:  (1) whether the submission is based on a company’s own previously cleared device or on a competitor’s device, and  (2) whether the submission includes a Predetermined Change Control Plan (PCCP).

As expected, submissions relying on a competitor’s device generally take longer to obtain clearance than those building on a company’s own prior device(median clearance time: 170 days vs. 120 days). This reflects the additional complexity involved in demonstrating substantial equivalence in less familiar or less directly comparable cases.

 

However, when comparing submissions with and without a PCCP, no meaningful difference in clearance time is observed. Forsubmissions using a company’s own predicate, the median review time is identical (120 days), regardless of whether a PCCP is included. The same overall pattern holds for submissions based on competitor predicates. While the median clearance time appears slightly lower for submissions with a PCCP (159 days vs. 170 days), this difference is most likely driven by the very small sample size (n=5) rather than a true effect.

Taken together, these results suggest that PCCPs do not meaningfully increase 510(k) review timelines. At the same time, they do not fully explain why PCCPs are predominantly introduced in follow-on submissions. Most likely, the “cost” of a PCCP is not primarily reflected in review time, but in the additional upfront work, uncertainty, and complexity during submission preparation—particularly for first-of-a-kind devices.

An important caveat is the presence of observation bias. This analysis only includes successfully cleared submissions and does not capture whether the inclusion of a PCCP affects the overall likelihood of clearance. In practice, if significant issues with a proposed PCCP arise during FDA review and cannot be resolved in a timely manner, manufacturers may choose to remove the PCCP from the submission. Such cases would not be visible in this dataset and could skew the observed results.

Does a PCCP actually remove the need for new 510(k)s?

The central promise of a PCCP is that it allows manufacturers to implement predefined changes without submitting a new 510(k). A key question, therefore, is whether PCCPs actually deliver on this promise in practice.

To explore this, we analyzed which 510(k) submissions that included a PCCP were later used as predicates for subsequent 510(k)s. In total, 25 PCCP-containing clearances were identified as predicates in follow-on submissions.

We then examined whether these follow-on submissions were submitted by the same manufacturer. In 19 of these cases, the manufacturer itself submitted a new 510(k), despite having previously included a PCCP intended to enable changes without requiring additional submissions. This suggests that, in practice, PCCPs do not fully eliminate the need for follow-on 510(k)s.

Understanding why PCCPs do not fully eliminate the need for follow-on 510(k)s requires a closer look at the types of changes being made. One of the key limitations is explicitly stated in FDA guidance:

“In general, FDA believes that modifications included in a PCCP should also maintain the device within the device’s indications for use. As with modifications to the intended use, FDA believes that most modifications to the indications for use included in a PCCP would be difficult for FDA to assess prospectively to determine whether the device would remain safe and effective.”

This limitation is clearly reflected in practice. A number of follow-on 510(k)s are driven by changes to the intended use or indications for use, changes that fall outside the scope of an approved PCCP.

For example, the Ceribell Infant Seizure Detection Software was originally cleared for patients aged 1 year and older (K241589). While the submission included a PCCP for algorithm updates, extending the use to patients below 1 year of age required a new 510(k) (K252070).

Similarly, Tyto Insights (an AI-based algorithm analyzing digital stethoscope signals) was initially cleared to identify cases suspected of “Rhonchi” (K240555), with a PCCP covering retraining, performance improvements, and adding compatible devices. However, expanding the indication to include identification of lung sounds suggestive of “Crackle” required a new 510(k) (K243567).

These examples illustrate a fundamental constraint: updates to intended use or indications for use are, by design, excluded from PCCPs and will trigger new submissions.

A second major limitation is more practical: PCCPs require manufacturers to anticipate future changes in advance. In reality, this is not always feasible.

For instance, MammoScreen BD, an AI-based system for analyzing mammograms, was originally cleared with a PCCP that anticipated compatibility expansions to Siemens and GE systems (K241561). However, when the manufacturer later sought to add compatibility with the Hologic Envision system, this fell outside the predefined scope and required a new 510(k) (K243685).

In other cases, new (software) features triggered additional submissions because they were not part of the original PCCP.

Medtronic, Inc. submitted a 510(k) for the Affera Integrated Mapping System that included a PCCP covering several anticipated changes (K241828). However, a subsequent update introducing a new calibration feature was not included in the original plan, necessitating a new 510(k) (K243892). As this was a relatively minor change, it was submitted as a Special 510(k) and cleared within 30 days.

A similar pattern can be seen with BrightHeart. Their BrightHeart View Classifier was initially cleared with a PCCP using a third-party predicate (K243684). This submission took 159 days and was cleared on May 7, 2025. Shortly after, the company submitted an update to their display method via a new Special 510(k), just five days after the initial clearance, which was approved in 24 days.

Together, these examples highlight both the limitations and the practical trade-offs of PCCPs. While they enable predefined, well-scoped updates, they cannot capture all future product evolution. At the same time, they illustrate a viable alternative pathway: when changes are sufficiently contained and can be verified using existing methods, a Special 510(k) can provide a fast and efficient route to clearance.

What does this mean for regulatory strategy? To PCCP or not PCCP…

The short answer: it depends.

More specifically, it depends on where you are in your product and company lifecycle and what you need your next FDA submission to achieve.

If you are pursuing your first FDA clearance, the priority is often clear: get to market. A first US clearance can be a critical inflection point, unlocking commercial traction, partnerships, or a new round of funding. At this stage, there may still be uncertainty around substantial equivalence, the level of evidence required, and how the FDA will view your device.

In this context, adding a PCCP may not always be the optimal choice. While our analysis shows that PCCPs do not measurably slow down review timelines, they do introduce additional complexity during submission preparation and review. If key aspects of the device (such as intended use, indications, or technical positioning) are still being actively discussed with the FDA, including a PCCP can increase the scope of those discussions.

A pragmatic approach, therefore, is often to focus the initial 510(k) on achieving clearance with the clearest and most robust case possible. A follow-on submission, using your own device as the predicate, can then build on that foundation. At that point, you have a much better understanding of FDA expectations, your validation framework is established, and you are in a stronger position to define meaningful and realistic PCCPs.

Importantly, early-stage teams should also consider flexibility. Product direction can evolve rapidly in the first years after clearance, and it is often difficult to predict which changes will matter most over a 1–3 year horizon. Over-specifying future changes in a PCCP too early can introduce constraints, while a follow-on 510(k) (particularly via a Special 510(k) where applicable) can be a fast and manageable way to implement updates. As shown in our examples, such submissions can be cleared in a matter of weeks when changes are well-bounded.

For more established organizations, the calculus can be different. Companies with a mature product line, clearer roadmaps, and longer planning horizons are often better positioned to benefit from PCCPs. In these settings, future updates—such as incremental algorithm improvements or planned compatibility expansions—can be defined with greater confidence at the time of submission. A well-scoped PCCP can then reduce the need for repeated submissions and create efficiencies over time.

At the same time, larger organizations often face longer internal development and submission cycles. In that context, incorporating a PCCP can help streamline downstream updates, even if it requires additional coordination upfront.

Across both scenarios, one theme is consistent: PCCPs are a strategic tool, not a default choice. They are most effective when the types of future changes are well understood, predictable, and aligned with the originally cleared indications and technology.

Ultimately, defining the right approach requires aligning regulatory strategy with product development, business milestones, and funding timelines. The question is not simply “can we include a PCCP?”, but rather “does including a PCCP help us move faster where it matters most?”

At MD squared B.V., we work with clients to answer exactly that question,combining detailed analysis of regulatory precedents with a deep understanding of product and business strategy. All this is aimed at giving our clients the relevant information from all angles to make informed choices. If you are exploring how PCCPs fit into your US regulatory approach, we would be happy to discuss how to structure the right path forward.