Beyond First Clearance: The Hidden Cost of Product Changes in the EU and US
We often focus on the first clearance. Getting market access for your product in either the EU or the US is a major hurdle and milestone, and a cause for much celebration. However, this should hopefully be the start of a journey of growth and continued learning to better serve your customers.
Naturally, this will require changes and updates to your product. But as discussed previously, this is where the fundamental difference between the US FDA and EU MDR regulatory regimes comes into play. In the US, the FDA reviews every technical file, and you have to submit any significant change for renewed approval.
In the EU, your QMS (including change management) is certified. This allows (most) changes to, or variants of, an existing product to be brought onto the market after completing your own internal verification and validation process according to your change procedures. The notified body does perform a technical documentation review before the initial device is brought to market. Subsequent changes are then checked during the yearly certification audits.
This difference in regulatory philosophy has a major impact on the regulatory overhead and delays associated with changes to existing devices. Getting this wrong will not only slow down innovation but can also result in enforcement actions from the FDA and other authorities.
In this article, we share a number of case studies based on publicly available information. None of the companies mentioned are our clients, and all analysis is based solely on publicly available information. To limit the discoverability of this post by people searching for these companies or products, the company names have been withheld from the main text, but all underlying reference information is linked.
One QMS certificate vs 13 510(k)s
A UK-based AI radiology company has brought two major products to market: one focused on stroke and other brain diseases, and another focused on lung imaging. Their first US clearance was obtained via a 510(k) and “provided viewing and analysis capabilities for imaging datasets acquired with CTA”. Over the years, they updated their product and added novel features, with each major update requiring a new 510(k). The figure below shows how the different 510(k)s are linked to each other. One can clearly see the different product lines focused on brain and lung imaging. In total, they submitted 13 510(k)s, with an average review time of 145 days.

It is interesting to look at the company’s QMS certificate. The notified body declares that the QMS is suitable for the products mentioned on the certificate. As discussed in a previous post, under the EU MDR, Class IIb (non-implantable) products are grouped together based on their EDMN code. In this case, the relevant code is Z11030592 – Diagnostic and Interventional Radiology Instruments – Medical Device Software.
It appears that all of the different product variants fall within this same broad category. This means that the notified body will assess a sampling of the technical files under this code during surveillance audits. The manufacturer can bring a new product within this category to market without prior technical file review by the notified body.
It is important to realize what this does and does not mean. The manufacturer still needs to verify and validate the performance of their product. They will need to perform a clinical evaluation (potentially including a clinical investigation) and build a technical file. However, in the EU they can bring the product to market after completing these activities. In the US, they will generally need to perform many of the same activities, but then they must submit the file to the FDA and wait, on average, several months before being able to market the product. This applies to every significant update to the product.
To reduce the need for a new 510(k) for each algorithm change, their most recent 510(k)includes a Predetermined Change Control Plan (PCCP) that allows the following changes to be made:
- Retraining the model weights using expanded datasets enriched for specific subtypes.
- Adjustment of probability thresholds for specific subtype classes to optimize performance.
Any changes outside of these predetermined changes will still require a new 510(k).
Learnings from FDA warning letters
We have performed an automated review of all available FDA warning letters. These provide deep insight into different companies and the types of issues the FDA uncovers. Reading them can sometimes feel a bit voyeuristic. Here we focus on situations where the FDA determines that a product is “adulterated under section 501(f)(1)(B) of the Act, 21 U.S.C. § 351(f)(1)(B).” This means the company does not have a valid PMA or 510(k) for the product it is bringing to market. In most cases, the company does have a 510(k), but it has made changes that the FDA determines are significant under its guidance and therefore require a new 510(k).
Looking at the different warning letters is instructive in understanding how the FDA interprets this requirement:
- Warning Letter 1: Extending compatibility of software to interact with input devices other than those with which the device was originally cleared. The FDA noted that this change may introduce patient risks because of the use of additional input devices that had not been validated for compatibility with the software. Even if the manufacturer had performed the necessary testing to demonstrate compatibility, the FDA still considered this a significant change.
- Warning Letter 2: Changing from luer connections to bayonet connections in a vascular ultrasound device. This was determined to be a significant change because it was intended to reduce the risk of device misconnections. So even though the purpose was to reduce risk, the FDA concluded that a new 510(k) was required.
- Warning Letter 3: The company brought an updated version of its cleared microcatheter to market. It had changed the material and increased the maximum length compared to the cleared device. The FDA noted that “The changes made to the patient-contacting materials and larger surface area in the [New] Microcatheter can expose the patient to new or different amounts of chemicals compared to the [old] Microcatheter, which could result in adverse blood or tissue reactions.” This is indeed true, and one would expect the company to have updated its biological evaluation. However, under FDA regulations, these changes would still require a new 510(k).
- Warning Letter 4: This case involved a company that had an original clearance for a spirometer device. It created a new variant that included Bluetooth connectivity and added home use to the indication. The company submitted a new 510(k), but after receiving a Request for Additional Information from the FDA, it did not respond in time and the submission was therefore deemed abandoned. During a subsequent inspection, the FDA discovered that the company was still marketing the new device in the US, putting it in obvious regulatory trouble.
Looking at these situations and explanations, one has to wonder how many companies and products are out there where similar changes have been made over time. We all have been in these discussion is this change really significant? Or can we just do a letter to file and start shipping the new version? The FDA inspects only about 1,500 of the 25,000 registered establishments, meaning it sees only a fraction of manufacturers in any given year.
Conclusion
These cases illustrate the stark differences between the regulatory regimes in the EU and the US. Over time, this will significantly impact regulatory costs and effort. It is not always clear which regime is “easier” or “harder”.
The EU regime comes with ongoing regulatory maintenance costs to cover notified body surveillance audits, but it does allow faster product updates and modifications. For a company that makes few or no changes, costs in the US are limited largely to annual establishment registration fees, making it easier to weather a slower commercial ramp-up. However, every significant change may require a new 510(k), along with the associated costs and delay in market entery.
As discussed previously, this significant step change in regulatory burden creates a strong incentive to document a rationale that a change is non-significant and effectively roll the dice. While the likelihood of an FDA inspection may not be particularly high, the consequences can be substantial. In many of the cases referenced above, it appears that the companies ultimately decided to withdraw their products from the US market.
Predetermined Change Control Plans can play a role in increasing the pace of innovation for devices on the US market, but as discussed in a previous article, they require careful consideration and do not solve all of these challenges.
At MD squared B.V., we work closely with our clients to explore not only the initial path to market access and acceptance but also long-term success. This requires open and informed discussions, enabling clients to make choices that align with their innovation, business, and funding strategies.